Abstract
To determine polyethylene deformation in well-functioning meniscal bearing PCL-retaining total knee prosthesis a quantitative stereophotogrammetric analysis was carried out over 5 years' follow-up. Polyethylene plastic deformation was obtained by tracking the intermarker distance variations of five tantalum beads inserted into the polyethylene prior to surgery in a known position. Plastic deformation occurred in the posteromedial part of the meniscal bearing. Radiostereometric analysis was a useful tool to monitor the load distribution in vivo on the polyethylene of this meniscal bearing prosthesis. This analysis revealed a very similar pattern of plastic deformation in the polyethylene insert of this prosthesis design to the wear pattern revealed with different methodologies in PCL-retaining fixed-bearing prosthesis.
Original language | English |
---|---|
Pages (from-to) | 142-147 |
Number of pages | 6 |
Journal | Knee Surgery, Sports Traumatology, Arthroscopy |
Volume | 16 |
Issue number | 2 |
DOIs | |
Publication status | Published - Feb 2008 |
Keywords
- Mobile-bearing
- Polyethylene deformation
- Polyethylene wear
- Roentgen stereophotogrammetric analysis
- TKR
ASJC Scopus subject areas
- Orthopedics and Sports Medicine
- Physical Therapy, Sports Therapy and Rehabilitation
- Surgery