Abstract
A multicenter study of a recently developed ELISA for the determination of prothrombin fragment F1+2 was performed in order to evaluate analytical and clinical aspects. Mean intra-assay and inter-assay reproducibility were found to be 11.0 and 12.6%, respectively. The measuring range covered by the calibration curve reaches from 0.04 to 10.0nM/l F1+2. Testing 133 healthy subjects a reference range of 0.37 to 1.11nM/l F1+2 (2.5-97.5 percentile) with a median of 0.66nM/l F1+2 was calculated. Minor difficulties with blood sampling (venous occlusion for 2 min) did not affect F1+2 plasma concentrations. Significantly increased F1+2 levels were measured in patients with leukemia (p
Original language | English |
---|---|
Pages (from-to) | 413-417 |
Number of pages | 5 |
Journal | Thrombosis and Haemostasis |
Volume | 68 |
Issue number | 4 |
Publication status | Published - 1992 |
ASJC Scopus subject areas
- Hematology