TY - JOUR
T1 - Multicenter evaluation of the new immunoassay method for TSH measurement using the automated DxI platform
AU - Dittadi, Ruggero
AU - Rizzardi, Sara
AU - Masotti, Silvia
AU - Prontera, Concetta
AU - Ripoli, Andrea
AU - Fortunato, Antonio
AU - Alfano, Antonio
AU - Carrozza, Cinzia
AU - Correale, Mario
AU - Gessoni, Gianluca
AU - Migliardi, Marco
AU - Zucchelli, Giancarlo
AU - Clerico, Aldo
AU - Italian Section of the European Ligand Assay Society (ELAS)
N1 - Copyright © 2017 Elsevier B.V. All rights reserved.
PY - 2017/5
Y1 - 2017/5
N2 - AIM OF THE STUDY: Recently, Beckman Coulter Diagnostics set up a new TSH immunoassay for the automated DxI platform. The aim of this study was to evaluate and compare the analytical performance and clinical results of this method with those of previous method.MATERIAL AND METHODS: A multicenter study (named TSH ELAS Study) was organized using 593 serum samples, collected from healthy subjects and patients with thyroid disorders, and 13 control samples, circulated in an External Quality Assessment (EQA) scheme.RESULTS: The values of LoB and LoD, and LoQ at 20% CV were 0.0004mIU/L, 0.001mIU/L and 0.0023mIU/L, respectively. Moreover, TSH concentrations >0.01mIU/L actually show imprecision values lower than 5% CV. This new TSH assay showed a systematic underestimation (on average of 6.25%) compared to old method, which is mainly due to larger differences between methods for samples with low TSH concentrations, related to the better analytical sensitivity of new compared to old method. In a reference population, including 279 apparently healthy adult subjects, Caucasian volunteers (mean age 43.6years, age 20-63years, 138 women and 141 males) the distribution of TSH concentrations was: mean (CI 95%) 1.694mIU/L (1.588-1.779), median 1.495mIU/L (1.412-1.588mIU/L), 97.5th percentile 3.707mIU/L.CONCLUSIONS: The new TSH immunoassay for DxI platform shows some relevant improvements compared to the previous one: use of the most recent WHO 3rd IRP 81/563 standard and monoclonal antibodies (instead of polyclonal antibodies of the old method), and better analytical sensitivities and reproducibility.
AB - AIM OF THE STUDY: Recently, Beckman Coulter Diagnostics set up a new TSH immunoassay for the automated DxI platform. The aim of this study was to evaluate and compare the analytical performance and clinical results of this method with those of previous method.MATERIAL AND METHODS: A multicenter study (named TSH ELAS Study) was organized using 593 serum samples, collected from healthy subjects and patients with thyroid disorders, and 13 control samples, circulated in an External Quality Assessment (EQA) scheme.RESULTS: The values of LoB and LoD, and LoQ at 20% CV were 0.0004mIU/L, 0.001mIU/L and 0.0023mIU/L, respectively. Moreover, TSH concentrations >0.01mIU/L actually show imprecision values lower than 5% CV. This new TSH assay showed a systematic underestimation (on average of 6.25%) compared to old method, which is mainly due to larger differences between methods for samples with low TSH concentrations, related to the better analytical sensitivity of new compared to old method. In a reference population, including 279 apparently healthy adult subjects, Caucasian volunteers (mean age 43.6years, age 20-63years, 138 women and 141 males) the distribution of TSH concentrations was: mean (CI 95%) 1.694mIU/L (1.588-1.779), median 1.495mIU/L (1.412-1.588mIU/L), 97.5th percentile 3.707mIU/L.CONCLUSIONS: The new TSH immunoassay for DxI platform shows some relevant improvements compared to the previous one: use of the most recent WHO 3rd IRP 81/563 standard and monoclonal antibodies (instead of polyclonal antibodies of the old method), and better analytical sensitivities and reproducibility.
KW - Journal Article
U2 - 10.1016/j.cca.2017.02.006
DO - 10.1016/j.cca.2017.02.006
M3 - Article
C2 - 28189596
SN - 0009-8981
VL - 468
SP - 105
EP - 110
JO - Clinica Chimica Acta
JF - Clinica Chimica Acta
ER -