TY - JOUR
T1 - Is a dual-sensor pacemaker appropriate in patients with sino-atrial disease? Results from the DUSISLOG study
AU - Padeletti, Luigi
AU - Pieragnoli, Paolo
AU - Di Biase, Luigi
AU - Colella, Andrea
AU - Landolina, Maurizio
AU - Moro, Eugenio
AU - Orazi, Serafino
AU - Vicentini, Alfredo
AU - Maglia, Giampiero
AU - Pensabene, Orazio
AU - Raciti, Giovanni
AU - Barold, S. Serge
PY - 2006/1
Y1 - 2006/1
N2 - Background: Rate-responsive pacemakers (PMs) are often supplied with accelerometer (XL) and minute ventilation (MV) sensors to provide a physiologic rate response according to patient needs. No information is available about the real benefit of dual-sensor rate-responsive pacing on the daily life of patients. Methods: DUSISLOG (Dual Sensor vs Single Sensor comparison using patient activity LOGbook) is a two-arm prospective, randomized, multicenter study that enrolled 105 patients who received a rate-responsive PM (Insignia®, Guidant Corp.). After 1 month of DDD pacing at 60 ppm lower rate, a single sensor (XL or MV, randomized) was activated for 3 months at the manufacturer's suggested nominal settings, followed by a 3-month period with dual sensors optimized with automatic response. During the last month of each period, the following data concerning patient physical activity were retrieved from PM diagnostics (Activity Log): mean percentage of physical activity, mean intensity of activity. Quality of life (QoL) scores and 6-minute walk test (WT) were also recorded. Results: Single-sensor rate-responsive pacing resulted in symptomatic benefit equally with XL and MV sensors while no additional benefit was found using dual sensor. In a subgroup analysis, patients (17%) with marked chronotropic incompetence and with 0% atrial sensing received benefits from single sensor with an additional advantage from sensor (QoL: +21 ± 14% P <0.05; WT: +17 ± 7% P <0.02). Conclusion: In most patients with rate-responsive devices, a single sensor is sufficient to achieve a satisfactory rate response. A dual sensor combination and optimization provides an additional benefit only in a selected population with an advanced atrial chronotropic disease.
AB - Background: Rate-responsive pacemakers (PMs) are often supplied with accelerometer (XL) and minute ventilation (MV) sensors to provide a physiologic rate response according to patient needs. No information is available about the real benefit of dual-sensor rate-responsive pacing on the daily life of patients. Methods: DUSISLOG (Dual Sensor vs Single Sensor comparison using patient activity LOGbook) is a two-arm prospective, randomized, multicenter study that enrolled 105 patients who received a rate-responsive PM (Insignia®, Guidant Corp.). After 1 month of DDD pacing at 60 ppm lower rate, a single sensor (XL or MV, randomized) was activated for 3 months at the manufacturer's suggested nominal settings, followed by a 3-month period with dual sensors optimized with automatic response. During the last month of each period, the following data concerning patient physical activity were retrieved from PM diagnostics (Activity Log): mean percentage of physical activity, mean intensity of activity. Quality of life (QoL) scores and 6-minute walk test (WT) were also recorded. Results: Single-sensor rate-responsive pacing resulted in symptomatic benefit equally with XL and MV sensors while no additional benefit was found using dual sensor. In a subgroup analysis, patients (17%) with marked chronotropic incompetence and with 0% atrial sensing received benefits from single sensor with an additional advantage from sensor (QoL: +21 ± 14% P <0.05; WT: +17 ± 7% P <0.02). Conclusion: In most patients with rate-responsive devices, a single sensor is sufficient to achieve a satisfactory rate response. A dual sensor combination and optimization provides an additional benefit only in a selected population with an advanced atrial chronotropic disease.
KW - Atrial chronotropic incompetence
KW - Cardiac pacemaker
KW - Cardiac pacing
KW - Rate-responsive pacing
KW - Sick sinus syndrome
UR - http://www.scopus.com/inward/record.url?scp=33645120136&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=33645120136&partnerID=8YFLogxK
U2 - 10.1111/j.1540-8159.2006.00301.x
DO - 10.1111/j.1540-8159.2006.00301.x
M3 - Article
C2 - 16441715
AN - SCOPUS:33645120136
SN - 0147-8389
VL - 29
SP - 34
EP - 40
JO - PACE - Pacing and Clinical Electrophysiology
JF - PACE - Pacing and Clinical Electrophysiology
IS - 1
ER -