A high-performance liquid chromatographic method for the determination of total plasma homocyst(e)ine [H(e)]after reduction with sodium tetrahydroborate and precolumn derivatization with o-phthaldialdehyde is described. The analyses, carried out on a reversed-phase C18 column, were based on spectrofluorimetric detection. The sensitivity was 1 pmol per injection and the intra- and inter-assay relative standard deviations were 1.8% and 5%, respectively. The plasma H(e) concentration determined in 40 healthy volunteers (20-60 years old) was 12.4±2.9 μM (mean±S.D.), in good agreement with reference values.

Original languageEnglish
Pages (from-to)171-176
Number of pages6
JournalJournal of Chromatography A
Volume593
Issue number1-2
DOIs
Publication statusPublished - Feb 28 1992