Enhanced dynamic wedge: Dosimetry and clinical implementation

Stefania Delle Canne, V. Landoni, T. Malatesta, R. Capparella, R. Fragomeni, L. Begnozzi

Research output: Contribution to journalArticlepeer-review

Abstract

The Varian Enhanced Dynamic Wedge is a dynamic modality implemented on Clinac 2100 C/D and 600 C linear accelerators in order to obtain wedge angles just by the sweeping of one of the secondary collimators (y1 or y2) across the field. The Segmented Treatment Tables of instruction define dose rate modification according to jaw position and the same tables are implemented on Cadplan treatment planning system. The EDW implementation supports standard wedge angles of 10°, 15°, 20°, 25°, 30°, 45° and 60°, and field dimensions up to 30 cm in the wedge direction. Before routine clinical use, dosimetric verifications are required to verify proper working of the modality and Cadplan calculation.

Original languageEnglish
Pages (from-to)289-293
Number of pages5
JournalPhysica Medica
Volume19
Issue number4
Publication statusPublished - Oct 2003

Keywords

  • Dosimetry
  • Enhanced Dynamic Wedge
  • Film
  • Treatment Planning

ASJC Scopus subject areas

  • Biophysics
  • Physics and Astronomy(all)

Fingerprint

Dive into the research topics of 'Enhanced dynamic wedge: Dosimetry and clinical implementation'. Together they form a unique fingerprint.