Controlli di accettazione di sistemi computed radiography

Translated title of the contribution: Acceptance testing of computed radiography systems

Lucia Riccardi, Maria Cristina Cauzzo, Roberto Fabbris

Research output: Contribution to journalArticlepeer-review

Abstract

Acceptance testing of computed radiography systems (CR) is required to verify not only image quality but also compliance with the manufacturer's specifications. Therefore, CR acceptance testing is manufacturer-specific. This paper describes a series of performance tests performed on a large number of CRs (FCR 5000, Fuji) recently acquired by our institution. In particular we describe the following tests: dark noise, uniformity, exposure calibration, linearity and autoranging, limiting resolution, noise and low-contrast resolution, spatial accuracy, laser-beam function, erasure thoroughness, aliasing, grid response. Special attention is given to the practical aspects related to measurement and subsequent image analysis. We report the results obtained in the various tests. No significant variations from the reference levels were found. Nonetheless, in some cases the operating procedures had to be adapted.

Translated title of the contributionAcceptance testing of computed radiography systems
Original languageItalian
Pages (from-to)676-688
Number of pages13
JournalRadiologia Medica
Volume110
Issue number5-6
Publication statusPublished - Nov 2005

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Acceptance testing of computed radiography systems'. Together they form a unique fingerprint.

Cite this