Abstract
The evaluation of hip abductor strength is useful in assessing of the outcome of hip surgery. Hand-held dynamometers are available, but they are less reliable in assessing hip abductor strength than some other muscle groups. We describe a new device designed to measure hip abductor strength, which is practical in a clinical setting. A system of constraints, pads and reference points was devised to make force measurements as little examiner-dependent as possible. Reproducibility was assessed in a controlled setting. The abductor strength of ten healthy young subjects (average age 28 years) was tested twice on each side by two independent examiners. Tests were performed in a supine position, eliminating the influence of gravity and examiner intervention. The results indicated high reproducibility, the maximal measurement uncertainty being within 1 N. Intra-class correlation coefficients ranged from 0.85-0.98 for intra-rater reproducibility, and 0.81-0.96 for inter-rater reporoducibility. The coefficient of variation was lower than 10%. The device described may be suitable for routine clinical assessment of patients after hip surgery.
Original language | English |
---|---|
Pages (from-to) | 512-517 |
Number of pages | 6 |
Journal | HIP International |
Volume | 20 |
Issue number | 4 |
Publication status | Published - Oct 2010 |
Keywords
- Dynamometer
- Hip abductors
- Muscle strength