TY - JOUR
T1 - A new nonfluoroscopic navigation system to guide pulmonary vein isolation
AU - Tondo, Claudio
AU - Mantica, Massimo
AU - Russo, Giovanni
AU - Karapatsoudi, Eleni
AU - Lucchina, Arianna
AU - Nigro, Franco
AU - Wild, Michaela
AU - Molinaro, Cristina
AU - Bavila, Roberto
PY - 2005/1
Y1 - 2005/1
N2 - Different techniques have been proposed to treat atrial fibrillation (AF) by catheter ablation. This study compares a new three-dimensional (3D) nonfluoroscopic navigation system with conventional fluoroscopy to guide pulmonary vein (PV) isolation. A total of 60 consecutive patients with paroxysmal or persistent AF were randomly assigned to 3D-guided ablation (group 1, n = 30), versus conventional fluoroscopy guidance ablation (group 2, n = 30). Complete PV isolation was achieved in both groups. The mean duration of fluoroscopy exposure (22 ± 8 vs 56 ± 10 minutes), and radiofrequency delivery (5 ± 1 vs 10 ± 3 minutes) were significantly shorter in group 1 than in group 2, P <0.05 for both comparisons). The mean procedural time in group 1 was longer (225 ± 15 minutes) than in group 2 (156 ± 10 minutes, P <0.05) due to the learning curve and time spent to generate the 3D maps. Over a mean follow-up of 7 ± 2 months, 6 patients (20%) in group 2 had AF recurrences compared to 3 patients (10%) in group 1 (ns). The new nonfluoroscopic 3D system allows a high-resolution reconstruction of the left atrium and PVs. It significantly reduces the mean radiofrequency delivery and fluoroscopy times as opposed to ablation performed under fluoroscopy guidance.
AB - Different techniques have been proposed to treat atrial fibrillation (AF) by catheter ablation. This study compares a new three-dimensional (3D) nonfluoroscopic navigation system with conventional fluoroscopy to guide pulmonary vein (PV) isolation. A total of 60 consecutive patients with paroxysmal or persistent AF were randomly assigned to 3D-guided ablation (group 1, n = 30), versus conventional fluoroscopy guidance ablation (group 2, n = 30). Complete PV isolation was achieved in both groups. The mean duration of fluoroscopy exposure (22 ± 8 vs 56 ± 10 minutes), and radiofrequency delivery (5 ± 1 vs 10 ± 3 minutes) were significantly shorter in group 1 than in group 2, P <0.05 for both comparisons). The mean procedural time in group 1 was longer (225 ± 15 minutes) than in group 2 (156 ± 10 minutes, P <0.05) due to the learning curve and time spent to generate the 3D maps. Over a mean follow-up of 7 ± 2 months, 6 patients (20%) in group 2 had AF recurrences compared to 3 patients (10%) in group 1 (ns). The new nonfluoroscopic 3D system allows a high-resolution reconstruction of the left atrium and PVs. It significantly reduces the mean radiofrequency delivery and fluoroscopy times as opposed to ablation performed under fluoroscopy guidance.
KW - Atrial fibrillation
KW - Catheter ablation
KW - Mapping
KW - Pulmonary vein isolation
UR - http://www.scopus.com/inward/record.url?scp=12844277465&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=12844277465&partnerID=8YFLogxK
U2 - 10.1111/j.1540-8159.2005.00047.x
DO - 10.1111/j.1540-8159.2005.00047.x
M3 - Article
C2 - 15683472
AN - SCOPUS:12844277465
SN - 0147-8389
VL - 28
JO - PACE - Pacing and Clinical Electrophysiology
JF - PACE - Pacing and Clinical Electrophysiology
IS - SUPPL. 1
ER -